389 Views

Describe the operation of the Gow-Mac gas density detector


arpita laha 22nd Nov, 2017
Answer (1)
Vipul Chaudhary 25th Sep, 2020

Hi there!

The gas density balance invented by Martin (1) is one of the very first chromatographic detectors, but practical difficulties in its construction have prevented until recently any widespread utilization of this detector.

The Gow-Mac balance, patented by Standard Oil Co. (Indiana, U.S.A.), is, on the other hand, of surprising simplicity and at the same time offers comparable performance.

The purpose of the present study is to define the optimum operating conditions of the Gow-Mac balance, mounted on a chromatograph of our own construction.

A systematic study of response as a function of the nature of the carrier gas and of flow rates shows that the highest sensitivities have been obtained with CO 2 , then with nitrogen, and finally with argon, using the largest possible ratio of Reference/Measuring flow rates, but with the Reference flow rate constant for a given carrier gas, i.e., 4 liter/hr. for CO 2 and 6 liter/hr. for nitrogen and argon.

The gas density balance gives the concentration in weight per cent of the different constituents of a mixture directly, if the peak areas are multiplied by a coefficient that depends upon the molecular masses of the solute and the carrier gas.

The precision of such direct quantitative measurements is satisfactory, and every mixture thus analyzed can serve as a standard, either for determining the coefficients of response for any other detector, or for use with an automatic process chromatograph operating with a bar graph recorder.

Since it avoids quantitative calibrations, the gas density balance is of certain importance in a control laboratory where rapidity of operation is of the utmost importance.

Thank You.

Related Questions

Amity University-Noida B.Tech...
Apply
Among top 100 Universities Globally in the Times Higher Education (THE) Interdisciplinary Science Rankings 2026
VIT - VITEEE 2026
Apply
National level exam conducted by VIT University, Vellore | Ranked #16 by NIRF for Engg. | NAAC A++ Accredited
Amity University-Noida M.Tech...
Apply
Among top 100 Universities Globally in the Times Higher Education (THE) Interdisciplinary Science Rankings 2026
Amrita University B.Tech 2026
Apply
Recognized as Institute of Eminence by Govt. of India | NAAC ‘A++’ Grade | Upto 75% Scholarships
UPES M.Tech Admissions 2026
Apply
Ranked #45 Among Universities in India by NIRF | 1950+ Students Placed 91% Placement, 800+ Recruiters
JIIT Online MBA
Apply
Apply for Online MBA from Jaypee Institute of Information Technology
View All Application Forms

Download the Careers360 App on your Android phone

Regular exam updates, QnA, Predictors, College Applications & E-books now on your Mobile

150M+ Students
30,000+ Colleges
500+ Exams
1500+ E-books